Flash Point Tester, Closed Cup Flash Point Tester, Oil Testing Instrument manufacturer / supplier in China, offering Pmcc Flash Point Tester-Flash Point Tester-Closed Cup Flash Point Tester, Automatic Proximate Analyzer (TGA) -Coal Testing Equipment, Automatic Calorimeter-Automatic Laboratory Instrument-Lab Equipment Calorimeter and so on.

(/ )

Supplier Homepage Product Other Test Instrument Pmcc Flash Point Tester-Flash Point Tester-Closed Cup Flash Point Tester

Pmcc Flash Point Tester-Flash Point Tester-Closed Cup Flash Point Tester

FOB Price: US $1-950 / Piece
Min. Order: 1 Piece
Min. Order FOB Price
1 Piece US $1-950/ Piece
Get Latest Price
Production Capacity: 5000PCS/Years
Transport Package: Paper/Wood Box
Payment Terms: L/C, T/T, Western Union, Paypal, Money Gram

Request a custom order and have something just for you!

Send Customized Request
Basic Info
  • Model NO.: TR-TC-261
  • Color: Mix Colour
  • Type: Oil&Petroleum
  • Trademark: TRYTE
  • Origin: China/Hongkong
  • Quality: Same as OEM
  • Usage: Oil&Petroleum
  • Application: Base Oil, Derivatives, Oil&Petroleum, Industrial Lubricant, Oil Product
  • Specification: AC 100-240V, 50Hz/60Hz
Product Description

Pmcc Flash Point Tester-Flash Point Tester-Closed Cup Flash Point Tester

Purpose and scope:

    The instrument is designed and made as per the National Standard of People's Republic of China GB/T261-2008 "Test Methods for Flash Point of Petroleum Products (Pensky-Martens Closed Cup Methods)". It is suitable to determine closed cup flash point of petroleum products having a flash point higher than 40 °C as per GB/T261-2008. Refer to ASTM D93.    

Specification and parameters:

    1. Power supply: AC 220V±10%, 50 Hz;

    2. Heating device:

    (1) The furnace is made of SiC materials; its power is 600 W;

    (2) Heating power is adjustable in the range of 0~600W;

    3. Sample temperature rising rate: 1~12°C per minute, manually controlled;

    4. Electric stirring device:

    (1) Stirring motor: 45TYZ Motor;

    (2) Driving device: by a soft shaft;

    (3) Stirring blade specification: 8×40 mm;

    5. Stirring speed: As per the requirement of GB/T261-2008;

    (1) 90~120 RPM for procedure A;

    (2) 250±10 RPM for procedure B;

    6. Standard oil cup: As per the requirement of GB/T261-2008;

    (1) Inner diameter: 50.7~50.8 mm;

    (2) Depth: 55.7~56.0 mm;

    (3) Depth of Marked line on the oil cup: 33.9~34.3 mm;

    (4) Oil sample cubage: about 70 ml;

    7. Ignition device (Flame application device): 

    (1) Ignition source: Coal gas (or other civilian combustible gas);

    (2) Aperture of ignition device: 0.8 mm;

    8. Thermometer: Rod type Mercury Thermometer, as per GB/T261-2008;

    (1) Temperature range: -5~110°C, Each scale division is 0.5 °C;

    (2) Temperature range: 20~150°C, Each scale division is 1.0 °C;

    (3) Temperature range: 90~370°C, Each scale division is 2°C;

    9. Ambient temperature: ≤35°C;

    10. Relative humidity: ≤85%;

    11. Total power consumption: not more than 650 W;

    12. Dimensions (m): 0.42*0.35*0.35

Pmcc Flash Point Tester-Flash Point Tester-Closed Cup Flash Point Tester
Pmcc Flash Point Tester-Flash Point Tester-Closed Cup Flash Point Tester

Pmcc Flash Point Tester-Flash Point Tester-Closed Cup Flash Point Tester
Send your message to this supplier
Avatar
Mr. Robin
Chief Technology Officer
*From:
To: Mr. Robin
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now