Atomic Absorption Spectrophotometer-Atomic Absorption Spectrometer Machine

Application: Laboratory Apparatus
Warranty: > 1Year
Detection Method: Atomic Absorption Spectrophotometer
Advantage: High Resolution
Principle: Atomic Absorption Spectrophotometer
Wavelengths: 190-900nm
Customization:
Gold Member Since 2016

Suppliers with verified business licenses

Manufacturer/Factory & Trading Company

Basic Info.

Model NO.
TR-TCS-02 Atomic Absorption Spectrophotometer
Dispersive Element
Grating Spectrometer
Light Number
Polychromator
Certification
RoHS, ISO9001, CE, SGS
Customized
Customized
Transport Package
Paper/Wood Box
Specification
AC 100-240V, 50Hz/60Hz
Trademark
TRYTE
Origin
China/Hongkong
Production Capacity
1500PCS/Year

Product Description

Atomic Absorption Spectrophotometer-Atomic Absorption Spectrometer Machine

Features:
Fully Automated
  • The flame and graphite furnace is integrated into one instrument. The change over from one technique to another is carried out by simple keystrokes within the software.
  • The instruments have a motorized 8 hollow cathode lamp turret which allows the automatic positioning and optimization of each hollow cathode lamp by the software.
  • The control of the gas flows for the fuel gas (C2H2) and the positioning of the burner is also carried out directly from the software thus allowing optimization of the instruments for the best analytical parameters for a selected analysis.
  • Two methods of background correction are available. The first utilizes a deuterium lamp and the second is the proven method of self reversal.
  • The location of the wavelength and peak selection is automatic and controlled from the software.
  • The spectral bandwidth is automated and is available with a choice of five slit sizes
  • The electronic parameters for the photomultiplier detector, the hollow cathode lamp current and the balancing of the absorbance and background energies are controlled from the software.
  • The ignition of the flame is computer controlled and the various safety interlocks offer a very safe operating system.
 
ADVANCED GRAPHITE FURNACE
  • The unique design of the graphite furnace reduces the chemical interference effects and memory effects by uniformly heating the graphite electrode.
  • The computer controlled heating program allows the user to select the best heating program for the analysis.
  • The optical temperature during the atomisation stage ensures the rapid heating and rapid analysis. This helps to extend the life of the graphite tube and enhances analytical accuracy.
  •  
  • PROVEN SAFETY FEATURES
  • The flame conditions are continuously monitored and should the flow rates change an audible alarm sounds.
  • The pressure of the support gas (oxidant) is monitored constantly. If the pressure changes then the flow of the fuel gas will be stopped and the flame will be safely extinguished.
  • A sensor monitors the level of liquid in the drain and will prevent ignition if too low. The flame will also be extinguished of the level of liquid in the drain changes significantly.
  • The argon pressure for the graphite furnace is constantly monitored and should it change the heating cycle for the graphite electrode will immediately cease and the graphite electrode will be de-energized.
  • Cooling water flow rates for the graphite furnace are also monitored for changes and should changes occur the heating program will cease.
  • If the graphite tube should facture during the heating program the heating will cease.
Elements that can be analyzed with the TR-TCS-02
Atomic Absorption Spectrophotometer-Atomic Absorption Spectrometer Machine

Specifications
Optic System
Wavelength Range: 190nm - 900nm
Monochromator: Czerny-Turner configuration
Spectral Bandwidth: 0.1nm, 0.2nm, 0.4nm, 1.0nm, 2.0nm (5 steps. with automatic changeover)
Wavelength Accuracy: 0.25nm
Wavelength Repeatability: 0.15nm
Baseline Stability: 0.005A/30 min
 

Flame analysis
Sensitivity (Cu): 0.03µg/ml /1%
Detection Limit (Cu): 0.006µg/ml
Burner Head: Titanium alloy burner
Nebulizer: High-efficiency glass nebulizer
Atomization Chamber: Corrosion-resistant material
Position Adjustment: Automatic changeover of flame and furnace
Automatic setting of optimum height for flame burner
Safety: Automatic ignition and of mixing air-acetylene gas with safety control
 
Graphite furnace analysis
Character Value (Cd): 0.5×10-12g
Detection Limit (Cd): 1.0×10-12g
Temperature Range: Ambient - 2650
Heating: Voltage feedback control when drying and ashing;
Optical temperature control when atomizing
Heating program: Up to 9 steps with choices of ramp, temperature increase and full-power heating
 
Background correction
Deuterium Lamp Background
Correction:
1.0 Abs Self-Reversal Background Correction: 3.0 Abs
 
Data processing
Analytical method: flame, graphite furnace and hydride
Determination method: calibration curves using 1st, 2nd and 3rd order of fit, standard addition method
Repetitions: 1-20 with calculations of average, SD and RSD
Result Printout: output of parameters, data, spectra and calibration curves
 
Mainframe
Light Source: 8 hollow cathode lamp turrets with 2 lamps simultaneously lit (one lamp pre-heated)
Power Supply: 110V/60Hz or 220V/50Hz three-phase AC
200W (mainframe) 5000W (graphite furnace)
Dimensions mainframe 110 cm × 50cm × 45cm
Graphite furnace 50cm × 50cm × 45cm

Atomic Absorption Spectrophotometer-Atomic Absorption Spectrometer Machine     Atomic Absorption Spectrophotometer-Atomic Absorption Spectrometer Machine
                     
[VARIAN]                                                                      [PE]

Packing:

Atomic Absorption Spectrophotometer-Atomic Absorption Spectrometer Machine

Company Information:

      As a diversified manufacturer of high-tech products, it was founded in 1999, and devoted to the products, sales and technical service for various industries and fields like Energy analysis instrument, Oil, coal, steel, and geology; medical analysis and application instruments, biochemistry analyzer, elisa reader, blood analyzer,and urine; as well as laboratory application instruments,autoclave/sterilizer,freeze dryer, incubator,drying oven,balance,ICP,DRS,HPLC,gas chromatography,and spectrophotometer,meanwhile,TRYTE Technologies integrated and purchased some companies of PCB board, highway instruments at home and abroad. Our company has become the "one stop" instrument procurement service provider worthy of the name.
      TRYTE technologies (TRYTE) mainly sell products by selling direct to customer, but also through distributors, resellers and manufacturer representatives, telephone sales and e-commerce sales, customers cover all over 80% of the world's countries and regions.TRYTE have solutions with rules adaptability according to the specific requirements of the design, from the instrument to some laboratories or departments.Whichever kind of solution you choose, you can fully trust it, because it is proved from 100, 00 successful results, leading development process, and decades of actual certification experience.
Atomic Absorption Spectrophotometer-Atomic Absorption Spectrometer Machine


Why choose us?

1. We supply high quality products with reasonable price.
2. We are the professional manufacturer;
3. Advanced production machine;
4. Strict quality control system;
5. Ten people research&develop team;
6. We have professional sales team;





 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Gold Member Since 2016

Suppliers with verified business licenses

Manufacturer/Factory & Trading Company
Number of Employees
30
Year of Establishment
2008-07-03