Auto Microplate Reader-Micro Plate Elisa Reader-Elisa Reader Instrument
Main features:
Big screen all-English display, English menus, touching screen, easy to operate;
Qualitative and quantitative processing system is fully functional, with a rich formula and curve equation, and can be customized settings;
8-channel optical fiber high speed scanning special center orientation for alisa hole leads to quick and correct measuring result;
Measurements: single wavelength, double wavelength, two points, kinetic method, multi-wavelength, external computer-controlled measurement, automatic identification of filters;
Calculating method: absorbance method, coefficient of concentration method, the standard concentration method, standard curve, single limit detection, double limit detection, level detection, column subtraction;
Reporting: raw data, qualitative, semi-quantitative and quantitative analysis of the result and quality control reports;
Large-screen all-Chinese display users' options and formula settings, can display 96 well plate information, with measurement results immediately displayed in full screen;
Open programming, with reagents and reaction plate optional;
With the function of self-test and self-calibration;
With built-in microtiter plate vibration for automatic control;
With built-in high-speed thermal printer;
Main technical parameters:
Wavelength range: 400-750nm
Interference filter: 8 filter positions (built-in), standard configurations of 405, 450, 492, 630 (also according to clients' requirement).
Measurement range: 0.000A~4.000A
Degree of linearity:±1%
Accuracy: ±1%
Absorbance repetition:≤0.5%
Stability :constantly stable due to automatically deduct background
Testing speed:5 seconds/96 holes (single wavelength),
10 seconds/96 holes (double wavelength)
Display: all-English LCD
External interface: RS-232 double-way communication interface
Storage: can store 200 testing procedures and 96-well plates 1000 original data, which are to be displayed, stored as well as printed
Resolution: 0.001 A
Input way:touching screen and pen
Details:
Display
|
240×128 LCD Touch screen(blue Backlight)
|
Computer Connector
|
RS232 9 needle
|
Suitable Micro plate Type
|
96 wells, Flat bottom, U bottom, V bottom
|
Shake
|
3 kind of intensity, can fit the exact time
|
Memory
|
200 programs, 1000 plates result
|
Light Source
|
Tungsten Halogen lamp, lamp saver function, 12V,20W
|
Wavelength range
|
400-750nm
|
Optical sensor
|
8 photodiode
|
Standard filters
|
405,450,492,630nm standard filter 4,
|
Optional filters
|
4free positions for optional filters
|
Reading speed
|
single wavelength<5sec/single, Duble wavelength<10sec
|
Absorbance range
|
0.000 - 3.500 Abs
|
Resolution
|
0.001 Abs
|
stability
|
±0.003 Abs
|
accuracy
|
ΔA≤±0.01(0≤A<1.0);
|
≤2.5%(A≥1.0);
|
|
Sensitivity
|
≥0.01 Abs
|
Repeatability
|
≤0.2% (0≤A≤1.5Abs);
|
≤1.0% (A>1.5Abs);
|
|
8 Channel differences
|
≤0.02
|
Linearity
|
≥0.999
|
Internal printer
|
57mm thermal printer
|
Voltage
|
AC 100-240V, 50Hz/60Hz
|
Power
|
90VA
|
Volume(Instrument)
|
318*286*168mm
|
Net weight
|
6.5Kg
|
Gross weight
|
8.7Kg
|
Outer box
|
45*41*29cm
|
Working temperature
|
10ºC-30ºC
|
Working humidity
|
≤ 85%
|
Storage temperature
|
0ºC-55ºC
|
Storage humidity
|
≤ 95%
|
Operation language:
|
English and Chinese
|
PC software
|
optional accessories
|