Customization: | Available |
---|---|
Brand: | Tryte |
Customized: | Non-Customized |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
1. Measures CO2, CH4, O2, H2S gases with customized measure range
2. Modular sensor design ensures easy on operation and maintenance
3. Self-developed gas conditioning device to remove vapor and dust in biogas for analyzer's protection in long term operation
4. Built in rechargeable lithium battery for easy carrying
5. Stores and downloads readings available.
Technical specifications | ||||||
Component | Method | Measuring range | Resolution | Precision | Repeatability | |
CH4 | NDIR | 0-100% | 0.01% | ≤ ± 2% FS | ≤ ±2% FS | |
CO2 | NDIR | 0-50% | 0.01% | ≤ ± 2% FS | ≤ 2% FS | |
O2 | ECD | 0-25% | 0.01% | ≤ ± 3% FS | ≤ 2% FS | |
H2S | ECD | 0-9999 ppm | 1 ppm | ≤ ± 3% FS | ≤ 2% FS | |
Note: Measuring range can be defined as customer request. | ||||||
Flow | 0.7-1.2L/min | |||||
Inlet pressure | 2kPa -50kPa | |||||
Sampling gas condition | Moisture removed, dust filtered (<1um) | |||||
Response time (TD+T90) | <10s (NDIR) | |||||
Storage | Up to 1500 sets of 6 data | |||||
Communication interface | RS232 ( real time and memory data download software included) | |||||
Display | LCD 320 x 240 display with back-lit function | |||||
Working environment | ||||||
Work temperature | 0-50ºC | |||||
Relative humidity | ≤ 95% | |||||
Ambient pressure | 86 - 108 kPa | |||||
Working power supply | External 220 VAC-50Hz, Internal with battery and charger; Autonomy of > 4h with pump in operation | |||||
Dimension | 380 × 140 × 255mm (Width*length*height) | |||||
Casing | Casing: Robust casing in aluminium with cover and shoulder trap | |||||
Weight | 5 kg max |
Instruction:
Gasboard-3200L portable biogas analyzer is designed to measure CH4, CO2, H2S and O2 concentration simultaneously in different biogas applications, like landfill, water treatment, CDM projects, anaerobic digestion and other fermentation processes, based on NDIR technology for CH4, CO2 gases, ECD technology for H2S and O2 gases. This model can be powered by lithium battery ensure easy carrying among different measuring points
Packing: